The TV2 is a highly engineered monitor/alarm/logger with a simple user interface. It uses
remote wire or wireless sensors to collect temperature, humidity and/or pressure data. The
Easy-Touch™display shows current conditions as well as days, weeks, months and even years of
logged data. It complies with all federal regulations.
It displays current, maximum /minimum conditions and digital charts for four remote sensors.
It logs and charts all monitored conditions which it stores in non-volatile memory for
protection. All stored data is encrypted to comply with 21 CFR 11, and can be copied to a
computer for archiving or printing through a USB cable. All captured data can be viewed in
graphic form or exported to an Excel file.
The TV2 stores up to 80,000 data points for each sensor, which means that it will store over
1.5 year of temperature history if you are sampling temperature once every 10 minutes. If
you are logging temperature and humidity it stores nine months of history for each of the
four sensors. Although the TV2 can store and display temperature from absolute zero to
thousands of degrees, the actual temperatures collected depends on the type of sensors being
used (see below).
In addition to showing the current conditions the TV2 can display the logged data for each
sensor. Touching one of the displayed values on the QuickCheck display causes the logged
data to be displayed as an easy-to-read chart. The chart can be scrolled backwards and
forward. It can also be zoomed into or out of to show more or less data.
Technical specifications and performance details for the TV2 Refrigerator-Freezer Alarm System.
| Channels |
Temperature only sensors- Four Channels
or Temperature/humidity sensors - Eight Channels (4 temp & 4 RH) Pressure sensors - Four channels |
|---|---|
| Current Conditions Displayed |
1. One, two, three or four, sensors (Temp, Humidity or pressure) 3. Low backup battery icon 4. Max/Min value of each sensor 5. Alarm enabled/disabled for each sensor 6. Sensor battery icon (for wireless sensors only) 7. Date and time 8. Last time data reviewed |
| Display Resolution | 0.1° C/F, 1.0% RH, 0.000" or hPA LCD, 0.0001 on PC |
| Monitoring | Continuous with active display (Updating every 4 seconds for wired sensors. Wireless sensors update every 20 seconds) |
| Alerts | Audible and visual indication on Display Email: uses local internet connection (requires separate dialer) |
| Alarm Settings | High temperature High Humidity Low Temperature Low Humidity High Pressure Low Pressure Power out Low battery Missing sensor for wireless sensors |
| Max/Min Display | Shows on active display. Resets by touch. |
| Accuracy & Range | Sensor dependent |
| Size of TV2 Monitor | 8 3/8" x 5" x 1 3/8" |
| logging Interval | User selectable: 1/min - 1/24hrs |
Technical specifications and performance details for the TV2 Refrigerator-Freezer Alarm System.
| Thermistor sensor | -30°C to 75°C (±0.3°C), ( ±0.2°C option) submersible |
|---|---|
| Thermistor/Humidity sensor |
-30°C to 75°C (±0.3°C) 0% to 100% RH (±3.0%) w ±0.5% Repeatability |
| 4 Sensors (Wired/Wireless) | Extreme Temperature Wireless Sensors |
| K Thermocouple sensors* | -200°C to 1250°C. Accuracy ±2.2°C or ±0.75%, below 0°C ±2% |
| E Thermocouple sensors* | -200°C to 900°C. Accuracy ±1.7°C or ±0.5%, below 0°C ±1% |
| J Thermocouple sensors* | -0°C to 750°C. Accuracy ±2.2°C or ±0.75% |
| T Thermocouple sensors* | -250°C to 350°C. Accuracy ±1.0°C or ±0.75%, below 0°C ±1.5% |
Technical specifications and performance details for the TV2 Refrigerator-Freezer Alarm System.
| Digital temperature sensor | -30°C to 80°C (±0.2°C) submersible ±0.005°C repeatability |
|---|---|
| Thermistor sensor | -20°C to 75°C (±0.3°C), ( ±0.2°C option) submersible |
| Thermistor/Humidity sensor |
-20°C to 75°C (±0.3°C) 0% to 100% RH (±3.0%) 0.5% Repeatability |
| Extreme Temperature Wired Sensors | |
| K Thermocouple sensors* | -200°C to 1250°C. Accuracy ±2.2°C or ±0.75%, below 0°C ±2% |
| E Thermocouple sensors* | -200°C to 900°C. Accuracy ±1.7°C or ±0.5%, below 0°C ±1% |
| J Thermocouple sensors* | -0°C to 750°C. Accuracy ±2.2°C or ±0.75% |
| T Thermocouple sensors* | -250°C to 350°C. Accuracy ±1.0°C or ±0.75%, below 0°C ±1.5% |
| Differential Pressure Sensors for Positve or Negative Pressurized Areas | |
| Pressure Sensor | ±0.500"wg (Accuracy ±0.002"wg) |
Technical specifications and performance details for the TV2 Refrigerator-Freezer Alarm System.
| Calibration | Optional NIST traceable (1,2 or 3 points) |
|---|---|
| Characterization | Three point calibration table for each sensor |
| Controls | Easy-Touch™display responds to a finger touch |
| Data Storage Technique | Non-volatile memory storage of 80,000 samples per sensor |
| Power | Via USB cable or optional power adaptor (90 to 240vac) |
| Battery Backup | ~100 hrs while in sleep mode. 1100mAh rechargable lithium battery. Not user serviceable. |
| Output to PC | 6' USB cable (Note: if USB cable is plugged into PC, power is supplied by PC) |
| PC Software | Free site license included with Email alert |
| Indicators | LCD & LED (red/green) |
| User Settings | On board password-protected menu |
| LCD Pixels | 488 horizontal x 270 vertical |
| LCD Temperature Range | Operating 0°C to 50°C, storage -20°C to 68°C |
| LCD size | 2.5" x 4.5" |
| LCD color display life | ~4.5 years |
| LED | Green indicates AC power present. Red - operating on battery. |
| Relay | Dry N/O with 2 terminal clamp <30v.> Fused for .5 amp |
| Mounting | Wall mount - Velcro. |
| Flush Mount | 6.5" x 8" or 7.5" x 9" - four holes for molly bolts or screws |
| Instrument Case Mount | 12" x 12" x 4" with knockout ports |
| Display weight | 1 lb |
| Shipping weight | ~3 lbs with power adaptor, sensors & cables |